Error Codes
IFW Serial Protocol Error Codes
ER=1 Wheel rotation exceeds 2500 steps
ER=2 SBIG pulse is not within range (Astronomy models only)
ER=3 Invalid Wheel ID received
ER=4 Wheel is stuck in position
ER=5 Improper position received from PC
ER=6 Position change exceeds 800 steps
ER=7 8-position command for 5-position wheel
ER=8 12v power is disconnected (IFW 2 only)
HSFW HID Error Codes
No error has occurred. (cleared state)
The 12VDC power has been disconnected from the device.
The device stalled during a Home of Move procedure.
An invalid parameter was received during communication with the host(PC).
An attempt to Home the device was made while the device was already in motion.
An attempt to change filter positions was made while the device was already in motion.
An attempt to change filter positions was made before the device had been homed.
Optec use only
Unknown critical error